Product Category
Medical titanium wire
Use: Orthopedic implants
Specifications: Φ1.0-Φ7.0
Tolerance: h6-h9
Length: 2000mm or 3000mm fixed length/as per customer requirements
Implementation standard: GB/T13810-2017.ISO 5832-3 ASTM F136
Material properties: High material strength, large torsion angle, can meet the various needs of implant screws.
Chemical composition of medical titanium material
| N | C | H | Fe | O | Al | V | Pa | Mo | Ni | Ti |
Gr 1 | 0.03 | 0.08 | 0.015 | 0.20 | 0.18 | / | / | / | / | / | bal |
Gr 2 | 0.03 | 0.08 | 0.015 | 0.30 | 0.25 | / | / | / | / | / | bal |
Gr 3 | 0.05 | 0.08 | 0.015 | 0.30 | 0.35 | / | / | / | / | / | bal |
Gr 4 | 0.05 | 0.08 | 0.015 | 0.50 | 0.40 | / | / | / | / | / | |
Gr 5 | 0.05 | 0.08 | 0.015 | 0.40 | 0.20 | 5.5~6.75 | 3.5~4.5 | / | / | / | bal |
Gr 7 | 0.03 | 0.08 | 0.015 | 0.30 | 0.25 | / | / | 0.12~0.25 | / | / | bal |
Gr 9 | 0.03 | 0.08 | 0.015 | 0.25 | 0.15 | 2.5~3.5 | 2.0~3.0 | / | / | / | bal |
Gr12 | 0.03 | 0.08 | 0.015 | 0.30 | 0.25 | / | / | / | 0.2~0.4 | 0.6~0.9 | bal |
Performance of medical titanium material
grade | Tensil strength(min) | Yeild strength(min) | Elongation(%) | ||
ksi | MPa | ksi | MPa | ||
1 | 35 | 240 | 20 | 138 | 24 |
2 | 50 | 345 | 40 | 275 | 20 |
3 | 65 | 450 | 55 | 380 | 18 |
4 | 80 | 550 | 70 | 483 | 15 |
5 | 130 | 895 | 120 | 828 | 10 |
7 | 50 | 345 | 40 | 275 | 20 |
9 | 90 | 620 | 70 | 438 | 15 |
12 | 70 | 438 | 50 | 345 | 18 |
Previous
Related Products
Product Inquiry
Please leave your correct information, and we will contact you as soon as possible!
Contact Us
No. South Section, Taicheng Road, Weibing District, Baoji City, Shaanxi Province
Social Media Sharing
Online Message
Copyright ©Shaanxi China Titanium New Materials Co.,Ltd. All rights reserved.